Electric Analytical Balance Calibration, Cleaning & Daily Accuracy Check

Electric Analytical Balance Calibration; Purpose : Electric Analytical Balance Calibration; The purpose of this SOP is to describe the operation, calibration, cleaning and daily accuracy check of Electric Analytical Balance Scope : This procedure is applicable for Electric Analytical Balance , installed in the quality control laboratory of XX Pharmaceuticals Limited. Definitions / Abbreviation: [][]SOP: […]


Electric Analytical Balance Calibration; Purpose :

Electric Analytical Balance Calibration; The purpose of this SOP is to describe the operation, calibration, cleaning and daily accuracy check of Electric Analytical Balance

Scope :

This procedure is applicable for Electric Analytical Balance , installed in the quality control laboratory of XX Pharmaceuticals Limited.

Definitions / Abbreviation:

[][]SOP: Standard Operating Procedure
[][]QC: Quality Control

Responsibilities:

Executive/Sr. Executive, QC

[][]To ensure that this procedure is followed.
[][]To ensure daily accuracy check of the balances.
[][]To maintain the records properly as per SOP.
[][]To preserve calibration, daily accuracy check record.

Manager, Quality Control

[][]To ensure that this procedure is kept up to date.
[][]To confirm that the SOP is technically sound and reflects the required working practices.
[][]Arrange training on the SOP to all concerned personnel.
[][]To ensure implementation of the SOP after training.

Head of Quality Assurance

[][]Approval of the SOP
[][]To ensure the overall implementation of the SOP.

Procedure:

General precautions:

[][]Do not move or replace Balance after calibration. If require, calibrate after movement.
[][]Do not place Balance near to heat generating equipment or direct sunlight.
[][]Clean up any spillage in or around the balance immediately after operations.
[][]Proper protective clothing such as safety glasses, gloves and Laboratory coat must be worn during weighing process.
[][]Do not use sharply pointed objects to operate the keyboard of balance.
[][]Do not use cleaning agents which contain solvents or abrasive ingredients.

Operation:

To start up:

[][]Level and adjust at each time it is moved to a new location. The balance is exactly horizontal when the air bubble is in the middle of the level glass. Adjust the two front leveling feet appropriately until the air bubble comes to rest exactly in the middle of the glass.
[][]Plug the AC adapter and connect the power supply. Connect the printer to the balance.
[][]Switch on the printer using I/O switch.
[][]Press On/Off () button to on the Electric Analytical Balance. Display will show 0.0000g.
[][]Convert the weighing unit from g into mg, press scroll down button if necessary.
[][]Allow the balance to warm up to enable it to adapt itself to the ambient condition.
[][]Long press for menu button and select ‘BASIC’ by single press Enter button
[][]Set date, time and unit using keys and Enter button and save the settings using button.
[][]Press cancel (C) button to return come out of that ‘BASIC’ option and return to its main menu

Weighing:

[][]Check the machine is clean and calibrated. Then remove all loads from weighing pan.
[][]When zero (0.0000 g) reading is displayed, the balance is ready for operation. If the display shows greater than or less than zero, press tare button to tare the balance.
[][]If the balance is not in the weighing mode, press and hold the key down until ‘WEIGHING’ in the display.
[][]Use handle/coupling element for the operation of the draft shield door of the both side of the instrument.
[][]Open right/left/upper sliding window of the instrument.
[][]Place empty container/weighing paper on the pan of the balance. Close the sliding window.
[][]The weight is displayed. Allow the reading to stabilize and wait until the instability detector disappears from the left side of the display and the stability beep sounds.
[][]Press Tare button to tare the balance. When zero (0.0000 g) reading is displayed, the balance is ready for weighing.
[][]The weighing-in aid is a dynamic graphic indicator which shows the used amount of the total weighing range. It can be recognized at a glance when the load on the balance approaches the maximum load.
[][]Add weighing sample (Maximum 120 g) on the container/weighing paper using weighing spoon. Close the sliding window. The net weight is now displayed.
[][]Allow the reading to stabilize and wait until the instability detector (O) disappears from the left side of the display and the stability beep sounds
[][]To get printout of the weight, press print button.
[][]For performing another weighing press Tare button to tare the balance and follow previous steps from to increase or decrease digit, press display resolution button.
[][]Finally, press and hold On/Off button button untill ‘SHUT OFF’ shows on the display.
[][]Fill up the Operation Log book for Semi-Micro Balance (as per Annexure-II).

Daily Accuracy Check:

[][]Perform the Internal and External Calibration as per below mentioned procedure
[][]Take 20 g, 70 g of certified weight and clean it with a soft cloth.
[][]Use tare button to tare the balance before start with a weighing.
[][]Always press button to tare the balance between the interval of one weight to another and observe whether the instability detector (O) disappears from the left side of the display or not and the balance will give a stability beep sounds.
[][]Open one sliding window and place the weights separately at the center of the weighing pan, close the sliding window and allow the reading to be stabilized and wait until the instability detector (O) disappear from the left side of the display and the stability beep sounds.
[][]Print the results by pressing print button if necessary.
[][]Record the value of those weights in Log Book for Daily Accuracy Check (as per Annexure-I).

Calibration:

Internal Calibration:
Manual Adjustment with Internal weight:

[][]Long press menu button and press scroll down button to go to ‘ADVANCED’ option.
[][]Select ‘ADVANCED’ by pressing Enter button.
[][]Press scroll down button once to select ‘CAL’ option by pressing Enter button.
[][]Then select ‘ADJ. INT’ using scroll down button and then press Enter button.
[][]Press Cancel (C) twice and save the settings if required by scrolling down button and press Enter button and then return to its main menu.
[][]Unload weighing pan Press and hold calibration button to execute ‘Internal Adjustment’.
[][]The balance adjusts itself automatically. The adjusting is finished when the message “ADJ.DONE” appears briefly on the display. The balance returns to the last active application and is ready for operation.
[][]After adjustment, the result of the calibration will be printed out automatically or press print (button for print copy.

External Calibration:

Manual Adjustment with External weight:

[][]Long press menu button and press scroll down button to go to ‘ADVANCED’ option.
[][]Select ‘ADVANCED’ by pressing Enter button.
[][]Press scroll down button once to select ‘CAL’ option by pressing Enter button.
[][]Select ‘ADJ. EXT’ by using scroll down and then press Enter button.
[][]Set the adjustment weight (100g) by using -/+ key for External Calibration and press Enter button.
[][]Then press cancel (C) button twice and save the settings if required by using -/+ key and press Enter button and then return to its main menu.
[][]Unload weighing pan Press and hold calibration button to execute ‘External Adjustment’.
[][]The required predefined adjustment weight value flashes on the display.
[][]Place adjustment certified weight on the center of pan. The balance adjusts itself automatically.
[][]When ‘0.0000 g’ flashes, remove the adjustment weight.
[][]The adjustment is finished when the message ‘ADJ DONE’ appears briefly on the display. The balance returns to the last active application and is ready for operation.
[][]After adjustment, the result of the calibration will be printed out automatically or press print button for print copy.

Eccentricity:

[][]Take 20 g, 70 g of certified weight and clean it with a soft cloth.
[][]Use tare button to tare the balance before start with a weighing.
[][]Always press button to tare the balance between the interval of one weight to another and observe whether the instability detector (O) disappears from the left side of the display or not and the balance will give a stability beep sounds.
[][]Place 50g weight at center, Left Rear, Right Rear, right Front, Left Front and center on the pan again separately.
[][]Allow the reading to be stabilized in each time while weighing and wait until the instability detector (O) disappear from the left side of the display and the stability beep sounds.
[][]Press print button separately to print each individual reading in same print page.

Linearity:

[][]Use Tare button to tare the balance before start with a weighing.
[][]Always press button to tare the balance between the interval of one weight to another and observe whether the instability detector (O) disappears from the left side of the display or not and the balance will give a stability beep sounds.
[][]Place 10 g, 20 g and 50 g on the center of pan separately.
[][]Allow the reading to be stabilized in each time while weighing and wait until the instability detector (O) disappear from the left side of the display and the stability beep sounds.
[][]Press Print ( ) button separately to print each individual reading in same print page.

Sensitivity:

[][]Use Tare button to tare the balance before start with a weighing.
[][]Always press button to tare the balance between the interval of one weight to another and observe whether the instability detector (O) disappears from the left side of the display or not and the balance will give a stability beep sounds.
[][]Place 100 g on the center of pan.
[][]Allow the reading to be stabilized in each time while weighing and wait until the instability detector (O) disappear from the left side of the display and the stability beep sounds.
[][]Press Print ( ) button separately to print each individual reading in same print page.

Repeatability:

[][]Use Tare button to tare the balance before start with a weighing.
[][]Always press button to tare the balance between the interval of one weight to another and observe whether the instability detector (O) disappears from the left side of the display or not and the balance will give a stability beep sounds.
[][]Place 100 g on the center of the pan for ten times separately.
[][]Allow the reading to be stabilized in each time while weighing and wait until the instability detector (O) disappear from the left side of the display and the stability beep sounds.
[][]Press Print ( ) button separately to print each individual reading in same print page.

Result:

[][]Record all the values for calibration in Calibration Information Sheet for Electric Analytical Balance (as per Annexure-III).
[][]Perform the Calibration of the Electric Analytical Balance once after every six months.

Cleaning:

[][]Ensure that no liquid comes into contact with the balance and AC adapter.
[][]Remove all objects from the balance.
[][]Clean the pan with soft brush followed by with soft clean cloth.
[][]Clean all of the parts of the balance with clean dry cloth.
[][]Set the parts of weighing pan carefully.

 

Annexure:

Annexure-I: Log book for Daily Accuracy Check of Electric Analytical Balance
Annexure-II: Operation Log book for Electric Analytical Balance
Annexure-II: Calibration Information Sheet for Electric Analytical Balance