SOPs

Electromagnetic Sieve Shaker Calibration with Cleaning & Operation

Electromagnetic Sieve Shaker Calibration, Purpose:

Electromagnetic Sieve Shaker Calibration, The purpose of this SOP is to describe the operation, calibration and cleaning of Electromagnetic sieve shaker (Model: Electrolab EMS-8) used for the sieve analysis in the quality control laboratory at XX Pharmaceuticals Ltd.

Electromagnetic Sieve Shaker Calibration Scope:

This procedure is applicable to the Electromagnetic sieve shaker (Model: Electrolab EMS-8), used for conducting sieve analysis and particle size distribution using test sieves, in the quality control laboratory of XX Pharmaceuticals Limited.

Definitions / Abbreviation:

[][]SOP: Standard Operating Procedure.
[][]QC: Quality Control.
[][]ASTM: American Society for Testing and Material.

Responsibilities:

The roles and responsibility is as follows:

Executive, QC

[][]To ensure that this procedure is followed.
[][]To maintain the records properly as per SOP.

Manager, Quality Control

[][]To ensure that this procedure is kept up to date.
[][]To confirm that the SOP is technically sound and reflects the required working practices.
[][]To arrange training on the SOP to all concerned personnel and to ensure implementation of the SOP after training.
[][]Schedule calibration of the instrument at the defined intervals.

Head of Quality Assurance

[][]Approval of the SOP.
[][]To ensure the overall implementation of the SOP

Procedure:

Precaution(s):

[][]The equipment should not be set near to room heating equipment or direct sunlight.
[][]Use mask, gloves and safety goggles during handling the instrument.

Operation:

[][]Check the machine is clean and calibrated.
[][]Ensure that the working bench is clean.
[][]Switch on the power from its main and power on the instrument.
[][]Display indicates: ELECTRO – MAGNETIC
[][]SIEVE SHAKER EMS-8
[][]Ensure all the sieve diameter and pore size should be as per ASTM standard. Keep received certificate in appropriate place.
[][]Use certified sieve.
[][]Select the sieves required as per monograph/ procedure.
[][]Check the integrity of the sieves.
[][]Ensure the sieve numbers are matching with the ASTM number given on the sieves.
[][]Place the sieves in the descending order of sieve No. on the collecting pan.
[][]Set the sieves on the instrument, shaker ensures that the value of the amplitude is set at the lowest.
[][]Set the desired time (Maximum 99 minutes) by pressing the time key provided below the “TIME” display.
[][]Set the desired amplitude (power level 5 to 20) by pressing the power key provided below the “POWER” display.
[][]Select Continuous or intermittent mode by pressing “CONTNUOUS” or “INTERMITTENT” key respectively.
[][]Weigh the sample and evenly spread the same on the top sieve and close the sieve.
[][]Place the sieve sets on the base of the holder firmly and lift the assembly to touch the top of the holder.
[][]Tight the assembly and then press “START” key to start the operation.
[][]The shifter works during the time and with the power level programmed. In case of intermittent mode, the shifter will start vibrate at interval of 0.5 second.
[][]After completion of this process carefully remove the sieve set.
[][]Collect the samples retained on each sieve and on the collection pan and weigh the quantity of sample retained on each sieve.
[][]Calculate the percentage of the sample retained on each sieve and put the analysis data in the report sheet of electromagnetic sieve shaker.
[][]Check the integrity of the sieves again, if not ok label it as rejected and discard the same as scrap.

Cleaning of the Sieve:

[][]Do not use dissolvent or any hard material for cleaning.
[][]For sieves less than 500 micron use ultrasonic and detergent. The sieves should be gently cleaned by keeping it in ultrasonic for 4 to 5 minutes.
[][]Cleaning with soft brush followed by ultrasonic cleaning can be used for sieves having mesh size more than 500 micron.
[][]After cleaning sieves should be rinsed thoroughly and left to dry in an upright position.

Cleaning of the Sieve Shaker:

[][]Switch ‘OFF’ the instrument and switch ‘OFF’ from the mains.
[][]Clean the outer surface of the instrument with Isopropyl alcohol and dry it with tissue paper.
[][]Clean the surrounding areas of instrument with a cloth dampen in water

Calibration:

[][]Calibration frequency: once in a year and after every maintenance.
[][]Check the time with calibrated and certified clock at 10, 20, 50, 70 and 90 minutes. The tolerance should be ± 1 minute.
[][]After completing the calibration, maintain the calibration record of Sieve shaker as per Annexure-I.

Annexure:

Annexure–I: Calibration Information Sheet for Electromagnetic Sieve Shaker.
Annexure–II: Operation logbook for Electromagnetic Sieve Shaker.
Annexure–III: List of Sieves.

Electromagnetic Sieve Shaker Calibration with Cleaning & Operation Read More »

Karl Fischer Titrator Calibration with operation & cleaning

Karl Fischer Titrator Calibration, Purpose:

Karl Fischer Titrator Calibration, The purpose of this SOP is to describe the operation, standardization of titrants, calibration, cleaning and maintenance of Karl Fischer Titrator (Model: Mettler Toledo V20 Compact) used for the determination of the water content in the quality control laboratory at XX Pharmaceuticals Ltd.

Karl Fischer Titrator Calibration, Scope:

This procedure is applicable to the Karl Fischer Titrator, used for the determination of water content, in the analysis of raw materials, intermediate and formulated drug product, installed in the quality control laboratory of XX Pharmaceuticals Limited.

Definitions/Abbreviation:

[][]SOP: Standard Operating Procedure
[][]QC: Quality Control
[][]KF: Karl Fischer

Responsibilities:

The roles and responsibility is as follows:

Officer/Executive/Sr. Executive,

[][]Quality Control
[][]To ensure that this procedure is followed.
[][]To maintain the records properly as per SOP.

Manager, Quality Control

[][]To ensure that this procedure is kept up to date.
[][]To confirm that the SOP is technically sound and reflects the required working practices.
[][]To arrange training on the SOP to all concerned personnel and to ensure implementation of the SOP after training.
[][]Schedule calibration of the instrument at the defined intervals.

Head of Quality Assurance

[][]Approval of the SOP.
[][]To ensure the overall implementation of the SOP.

Procedure:

Precaution(s):

[][]The equipment should not be set near to room heating equipment or direct sunlight.
[][]Use mask, gloves and safety goggles during handling the Karl Fischer reagent and methanol under Fume Cupboard.
[][]Do not change stirring speed between the pre-titration and the titration.
[][]Always take some more volume of liquid sample by a syringe and then inject the desired volume.
[][]Clean and dry syringes and needles before use. Rinse the syringe with the sample solution itself 2 – 3 times and then take the test sample. This will remove the moisture from syringe if any.
[][]While testing various samples of which the moisture contents are roughly known, test the sample having lowest moisture first and then go on to the next in order to avoid contamination.
[][]If moisture content of samples are unknown and very small sample to obtain pilot reading and then adjust the size of sample accordingly. Keep next sample ready and if some time is lapsed between two titrations. Press START, get to the end point then add sample immediately. This will eliminate the error due to lapse of time.
[][]If fresh KF reagent is added to the old stock in KF reservoir, mix the solution in reservoir properly and recalibrate the reagent for new factor.
[][]Keep the titration cell tightly closed with stopper and only open for an instant to add sample fast.
[][]When renewing the reagents, clean the titration cell thoroughly with methanol (or another suitable solvent) and may be dried in an even at approximately 70⁰C.
[][]Adjust the filling rate of dispensing burette to avoid the introduction of air bubbles in the burette.

Operation:

[][]Check the cleanliness of the instrument.
[][]Check the calibration status of the Karl Fischer Titrator.
[][]Check the clarity and color of the titrant.
[][]Switch ON the instrument. The instrument always performs a self-test before the display of the Titrator name appears.

Functional description:

[][]The control panel consists of an integrated touch screen and four additional keys, which are located next to the touch-sensitive surface of the display.
[][]Reset, the current task is aborted Info, calls to the online help for the current contents of the current dialog.
[][]Home, always returns to the Home screen (the blue Home keys are located at the bottom right and left of the touch screen).

Rinse and fill burette:

[][]To fill the burette and tubes with titrant and remove any air bubbles from the system, the burette should be rinsed three times with titrant. To do that, use the manual operation “Rinse burette”. Press Home if the home screen is not already displayed.
[][]Select the Manual button to open the Manual operations dialog and press Burette button.
[][]In the Burette dialog, use the Rinse button to access the parameters for the rinse process.
[][]Select the relevant titrant.
[][]For the “Cycles” parameter, enter “3” for three rinse cycles.
[][]Press Start to start the rinse procedure.
[][]When all rinse procedures will be completed, use the ok button to return to the Rinse dialog. The burette should be filled and the hoses should be free of air bubbles.

Filling the titration beaker:

[][]To fill the titration beaker with solvent, the solvent manager must be attached to the solvent bottle.
[][]In the Pump dialog (Home>Manual>Pump), select the “Fill” value for the “Action” parameter.
[][]The “Reset counter” parameter will be activated.
[][]Press Start.
[][]Titration beaker will be filled with solvent.

Performing a volumetric KF titration:

[][]Fill dried Methanol in glass vessel.
[][]In the Home screen, use the Methods button and select the method type “KF vol”. In the subsequent dialog, select the “Mix Time” and enter required time.
Select start button.
[][]Start analysis dialog will be displayed. In this dialog select Number of sample and put “3” and then press Ok.
[][]Select Start button. The system will perform a pre-titration as a basis for a water-free solvent.
[][]As soon as the continually determined drift value falls below a regulated required value, the system automatically switches to Standby mode. In the Standby mode the Start Concentration button will be activated.
[][]Press the Start Concentration button, weigh about 10 mg of purified water as a standard and add into the titration vessel and then enter the sample weight.
[][]Press Sample ID and put the sample ID.
[][]Confirm with Ok. The analysis will be started automatically.
[][]Perform 3 samples analysis and press Result>Statistic>Print.
[][]Average water content and relative standard deviation (RSD) of 3 samples will obtain on the print out.
[][]Standard value of RSD is not more than 2%.
[][]Press Reset and select start button.
[][]Start analysis dialog will be displayed.
[][]When Start button will be selected in the Start analysis dialog, the system first always performs a pre-titration as a basis for a water-free solvent.
[][]As soon as the continually determined drift value falls below a regulated required value, the system automatically switches to Standby mode. In the Standby mode the Start sample button will be activated.
[][]Press the Start sample button, add the sample and then enter the sample weight.
[][]Press Sample ID and put the sample ID.
[][]Confirm with Ok. The analysis will be started automatically.
[][]After completing titration, the Result dialog will be displayed and printed automatically.

Calibration:

[][]Fill dried Methanol in glass vessel.
[][]Select start button.
[][]Start analysis dialog will be displayed. In this dialog select Number of sample and put “3” and then press Ok.
[][]Select Start button. The system will perform a pre-titration as a basis for a water-free solvent.
[][]As soon as the continually determined drift value falls below a regulated required value, the system automatically switches to Standby mode. In the Standby mode the Start Concentration button will be activated.
[][]Press the Start Concentration button, weigh about 100 mg of Di-sodium tartrate dihydrate as a standard and add into the titration vessel and then enter the sample weight.
[][]Press Sample ID and put the sample ID.
[][]Confirm with Ok. The analysis will be started automatically.
[][]Perform 3 samples analysis and press Result>Statistic>Print.
[][]Average water content and relative standard deviation (RSD) of 3 samples will obtain on the print out.
[][]Standard value of RSD is not more than 1%.
[][]Record the result in the “Calibration Information Sheet for Karl Fischer Titrator” as per Annexure-I.
[][]Affix the label “CALIBRATED” on the instrument if all test results are satisfactory. If the calibration result is out of limit or any discrepancy is during calibration, affix “UNDER Maintenance” label and inform the respective supplier.
[][]Calibrate the instrument once in every six month.

Cleaning:

[][]After completion of the testing, press Drain>Start button for removing the solution from the test compartment and wash with methanol which is used for water determination. Wipe the knob of sample compartment with tissue paper.
[][]Switch ‘OFF’ the instrument and switch ‘OFF’ the mains.
[][]If the drift of the KF solution remains too high, fill the drying tubes with fresh molecular sieve or regenerate molecular sieve in a drying oven at 105 – 150⁰C for at least 24 hrs. Before regeneration rinse the molecular sieve used on the waste bottle with purified water as it contain SO2.
[][]Collect the waste solution into waste beaker at the same time all waste to be thrown to avoid corrosion.
[][]Clean the outer surface of the instrument with Isopropyl alcohol and dry it with tissue paper.

Annexure

Annexure-I: Calibration Information Sheet for Karl Fischer Titrator.
Annexure-II: Operation logbook for Karl Fischer Titrator.

Karl Fischer Titrator Calibration with operation & cleaning Read More »

Calibration of Melting Point Apparatus with operation & cleaning

Calibration of Melting Point Apparatus , Purpose :

Calibration of  Melting Point Apparatus, The purpose of this SOP is to describe the operation, calibration and cleaning of Automatic Melting-point apparatus (Model: BUCHI M-565) used for the determination of the melting point of raw materials in the quality control laboratory at XX Pharmaceuticals Ltd.

Calibration of Melting Point Apparatus, Scope :

This procedure is applicable to the Automatic Melting-Point Apparatus (Model: BUCHI, M-565), installed in the quality control laboratory of XX Pharmaceuticals Limited.

Definitions / Abbreviation:

[][]SOP: Standard Operating Procedure.
[][]QC: Quality Control.
[][]Melting point or Melting range: Melting point or Melting range is defined as the point at which or those points of temperature within which, the solid coalesces and is completely melted. The optical properties of most substances change at this transition.

Responsibilities:

The roles and responsibility is as follows:

Sr. Executive/Executive, QC

[][]To ensure that this procedure is followed.
[][]To maintain the records properly as per SOP.
[][]To ensure cleaning of Melting point apparatus maintaining safety rules.

Manager, Quality Control

[][]To ensure that this procedure is kept up to date.
[][]To confirm that the SOP is technically sound and reflects the required working practices.
[][]Arrange training on the SOP to all concerned personnel.
[][]To ensure implementation of the SOP after training.
[][]Schedule calibration of the instrument at the defined intervals.

Head of Quality Assurance

[][]Approval of the SOP
[][]To ensure the overall implementation of the SOP.

Procedure:

Precautions:

[][]Laboratory coat and safety glasses (to protect eye from capillary tube) must be worn while handling the instrument.
[][]Always ensure that the cooling air outlet at the back of the instrument is not covered so that the air can flow.
[][]Use only the BUCHI sample tubes provided, as other tubes may not correctly fit in the sample holder. Always handle these glass tubes with care.
[][]Never touch the furnace or sample just removed from the furnace. The temperature of the furnace can reach 400°C.
[][]Care should be taken while inserting the capillary into its cavity or taking out from the cavity to avoid breaking, take it out vertically.
[][]Check always colour of the silicone oil. Normally it is colourless. If it observed as yellow or brown, discard the oil. And take fresh oil in that container.
[][]Disconnect the power supply before moving or cleaning of the instrument.
[][]When not in use, keep the Instrument in a safe place.
[][]Prior to use, user must ensure that equipment is calibrated.

Operation:

[][]Check the machine is clean and calibrated.
[][]Prepare the sample to make it powder, if required with an agate mortar.
[][]Fill the sample in capillary tube to a height of approximately 3 – 6 mm with the help of sample loader.
[][]Plug the instruments into an acceptable power source.
[][]Turn the instrument ON by using ‘I/O’ switch in front of the machine.
[][]Choose the Melting Point settings for melting point determination by rotary knob.
[][]Make sure that all samples are removed.
[][]To create a new melting parameter, press ‘Edit’.
[][]Select Start temperature with the rotary knob and press ‘Next’.
[][]Select Stop temperature with the rotary knob and press ‘Next’.
[][]Select Temperature gradient with the rotary and press ‘Save’.
[][]Press ‘START’ to start the determination.
[][]Turn the rotary knob to select a sample ID/Lot. Press ‘Enter’ after selecting each word/number by rotary knob.
[][]When machine reaches the set the temperature, a beep sound will be given and display will show ‘Insert sample & START’.
[][]Insert three capillary tubes containing sample and press ‘START’.
[][]To register additional temperature stamps manually, see the tube through the front window, press the ‘Set’ button for related sample. Each button can be pressed three times. Registered temperature stamps are displayed in green.
[][]After finish, machine will show the result.
[][]To run through the result, rotate the rotary knob.
[][]To replay the recorded melting process, press ‘Relay’.
[][]Press ‘End’ or ‘STOP’ to return to the idle screen.
[][]The result will be printed out automatically or by print option.
[][]Fill up the Log book for Automatic Melting point Apparatus ( as per Annexure-II)
[][]For creating a method:
[][]Press ‘Method’ to enter the method menu.
[][]Press Options-New-ok.
[][]Enter a method name and press ‘SAVE’.
[][]Adjust Start temperature, Stop temperature and Gradient temperature by using rotary knob. Use ‘Next’ button to move from one option to another. After setting the all parameter press ‘SAVE’.

Calibration:

[][]Calibrate the Automatic Melting Point Apparatus once in a year.
[][]Determine the melting point of 4-Nitotolune, Diphenyl acetic acid, Caffeine, potassium nitrate standard.
[][]To start calibration, turn the rotary knob to calibration. Press ‘START’. The instrument automatically heats to the start temperature of the first substance (4-Nitrotolune). Prepare at least six sample of each substance standard.
[][]Press ‘START’ to start the determination. Turn the rotary knob to select a lot/Sample ID and press ‘Enter’. Press ‘Save’ after giving Lot/Sample ID.
[][]Choose the correct Pharmacopoeia melting point value with the rotary knob. The Pharmacopoeia melting point is written on the certified of analysis. Press ‘SAVE’.
[][]Press ‘START’ to continue. Insert three samples and press ‘START’ to start the determination.
[][]After determination, press ‘START’ again to continue and wait for the beep sound and then insert three more samples and press ‘START’.
[][]After each run, the current result is indicated. After six results are obtained with in a standard deviation of ± 0.2°C. The process moves on to the next substance. The process identical for Diphenylacetic Acid, Caffeine and potassium Nitrate and save.
[][]After determination of the melting point of all four standards, compile the results and record the results in Calibration Information Sheet for Automatic Melting Point Apparatus (as per Annexure-I).

Cleaning:

Clean the instrument once in a week or in between when water becomes dirty.
Switch ‘OFF’ the instrument.
De-dust the instrument daily externally with a clean dry cotton cloth.
Remove the glass window from the heating block periodically and wipe it clean with alcohol or acetone. Replace the window with a new one if it will not come completely clean.
Once in a week remove adhered dust by wet mopping using detergent solution. Afterwards wipe the surface with a clean dry cloth to remove traces of detergent and moisture.

Annexure:

Annexure-I: Calibration Information Sheet for Automatic Melting Point Apparatus.
Annexure-I: Operation Log book for Automatic Melting Point Apparatus.

Calibration of Melting Point Apparatus with operation & cleaning Read More »

Calibration of Refractometer with Operation and Cleaning

Calibration of Refractometer, Purpose :

Calibration of Refractometer, The purpose of this SOP is to describe the operation, calibration and cleaning of Refractometer (Model: Rudolph J257) used for the determination of the refractive index of raw materials in the quality control laboratory at XX Pharmaceuticals Ltd.

Calibration of Refractometer, Scope :

This procedure is applicable to the Refractometer (Model: Rudolph J257), installed in the quality control laboratory of XX Pharmaceuticals Limited.

Definitions/Abbreviation:

[][]Standard Operating Procedure (SOP): A written authorized procedure, which gives instructions for performing operations.
[][]QC: Quality Control.
[][]Refractive index: The refractive index of a medium with reference to air is equal to the ratio of the sine of the angle of incidence of a beam of light in the air to the sign of the angle of refraction of the refracted beam in the given medium.

Responsibilities:

The roles and responsibility is as follows:

Officer/Executive/Sr. Executive, Quality Control

[][]To ensure that this procedure is followed.
[][]To maintain the records properly as per SOP.

Manager, Quality Control

[][]To ensure that this procedure is kept up to date.
[][]To confirm that the SOP is technically sound and reflects the required working practices.
[][]Arrange training on the SOP to all concerned personnel.
[][]To ensure implementation of the SOP after training.
[][]Schedule calibration of the instrument at the defined intervals.

Head of Quality Assurance

[][]Approval of the SOP.
[][]To ensure overall implementation of the SOP.

Procedure:

Precaution(s):

[][]Clean off samples with water and then with methanol or a suitable solvent as soon as the measurement is completed in a way to avoid possible scratching in the prism.
[][]Do not use the strong solvents such as N,N- Dimethylformamide or N,N-Dimethylacetamide, Cresol, Phenols and other tar acids.
[][]Never use any hard or metallic objects on the prism surface (where the sample to be measured is placed) as scratches quickly spoil the definition of the borderline.
[][]Disconnect the power supply before moving or cleaning of the instrument.
[][]When not in use, keep the instrument in a safe place.
[][]Prior to use, user must ensure that equipment is calibrated.

Operating procedure

[][]Check the machine is calibrated.
[][]Plug the instruments into an acceptable power source.
[][]Turn the instrument ON.
[][]Wait about 30 minutes to warm up the Refractometer.
[][]Carefully clean the polished surface of the fixed prism with water and then with methanol and dry before applying a sample.
[][]Place a few drops of sample on the polished face of the lower Measuring prism.
[][]Lower the sample cover.
[][]Select the option “measure when stable”.
[][]Press the “Read” button to obtain the sample reading.
[][]Take print out of the result.
[][]Record the operation in the logbook.

Calibration with Purified water, Carbon Tetra chloride and Toluene:

[][]Plug the instruments into an acceptable power source.
[][]Turn the instrument ON.
[][]Wait about 30 minutes to warm up the Refractometer.
[][]Carefully clean the polished surface of the fixed prism with water and then with methanol and dry before applying a sample.
[][]Determine the refractive index of the purified water, Toluene and Carbon Tetrachloride at 20°C.
[][]Place a few drops of sample on the polished face of the lower Measuring prism.
[][]Lower the sample cover.
[][]Select the option “measure when stable”.
[][]Press the “Read” button to obtain the sample reading.
[][]Take print out of the result.
[][]Compile the result & take approval for the calibration information sheet (Annexure-I).
[][]Affix the label “CALIBRATED” on the instrument if all test results are satisfactory.
[][]If the equipment found out of calibration limit affix “UNDER Maintenance” label and inform the respective supplier.
[][]Calibration frequency: Once in a year.

Cleaning procedure:

[][]After completion of the testing switch off the instrument and switch off the mains.
[][]Clean the prism surface with purified water or alcohol or xylene or a solvent which will remove sample residue.
[][]Finally clean with purified water.
[][]Wipe it with tissue paper.

Annexure:

Annexure-I: Calibration Information Sheet for Refractometer with Water, Carbon Tetrachloride and Toluene
Annexure-II: Logbook for Refractometer

Calibration of Refractometer with Operation and Cleaning Read More »

Calibration of Viscometer With Operation And Cleaning

Calibration of Viscometer, Purpose :

Calibration of Viscometer, The purpose of this SOP is to describe the operation, calibration and cleaning of viscometer (Model: Brookfield LVDV-II+Pro) used for the measurement of viscosity in the quality control laboratory at XX Pharmaceuticals Ltd.

Calibration of Viscometer, Scope :

This procedure is applicable to the Brookfield viscometer, installed in the quality control laboratory of XX Pharmaceuticals Limited.

Definitions/Abbreviation:

[][]Standard Operating Procedure (SOP): A written authorized procedure, which gives instructions for performing operations.
[][]Viscosity: Viscosity is measure of fluid’s resistance to flow.
[][]FSR: Full Scale Viscosity Range.
[][]cP: Centipoise

Responsibilities:

The roles and responsibility is as follows:

Officer/Executive/Sr. Executive, Quality Control

[][]To ensure that this procedure is followed.
[][]To maintain the records properly as per SOP.

Manager, Quality Control

[][]To ensure that this procedure is kept up to date.
[][]To confirm that the SOP is technically sound and reflects the required working practices.
[][]Arrange training on the SOP to all concerned personnel.
[][]To ensure implementation of the SOP after training.
[][]Schedule calibration of the instrument at the defined intervals.

Head of Quality Assurance

[][]Approval of the SOP.
[][]To ensure overall implementation of this SOP.

Procedure:

Precaution(s):

[][]To measure high viscosity, choose a small spindle or select slow speed.
[][]Remove spindle before cleaning.
[][]Prior to operation check the labeling of the viscometer. The label is adjusted using the two leveling screws on the base. Adjust so that the bubble label on the top of the instrument within the circle.
[][]Viscosity measurements should be accepted within the equivalent % of Torque Range from 10% to 100% for any combination of spindle/speed rotation.
[][]Viscosity measurements should be taken under laminar flow conditions, not under turbulent flow conditions.

Calibration procedure:

[][]Switch on the power from its main and power on the viscometer.
[][]Transfer the viscosity standard fluid (Standard Fluid 10, Standard Fluid 500 & Standard Fluid 5000) into 600 ml beaker.
[][]Immerse the beaker along with the spindle into water bath for 1 hour at set temperature 25.0°C.
[][]After 1 hour, check the temperature of the viscosity standard fluid with an accurate thermometer.
[][]Lower the DV-II+ into measurement position of the guard leg.
[][]Attach the spindle to the viscometer. In case of disk – shaped spindle, avoid trapping air bubbles beneath the disk by first immersing the spindle to an angle and then connecting to it to the viscometer.
[][]Measure the viscosity reading normally at 25.0°C ± 0.1°C, must rotate the spindle at least five times before readings are taken.
[][]The viscosity reading should be equal the cP value on the fluid standard to within the combined accuracies of the viscometer & viscosity standard i.e. ± (1% of full scale viscosity range + 1% viscosity Standard Fluid stated value at 25ºC).
[][]Calculate the Full scale viscosity is determined from following equation:
[][]Full scale viscosity [cP] = TK * SMC * 10000/ RPM
Where,
TK = spring torque constant of viscometer (for LVDV-II +Pro model TK value is 0.09373 )
SMC = spindle multiplier constant for spindle [SMC value for the four spindle LV1 (entry code: 61), LV2 (entry code: 62), LV3 (entry code: 63), LV4 (entry code: 64) and LV5 (entry code: 65) are respectively 6.4, 32, 128, 640 and 1280.] [][]RPM: Rotation per minute Compile the result & take approval for the calibration information sheet (Annexure-I).
[][]Carry out the test once in a year.
[][]Affix the label “CALIBRATED” on the instrument if all test results are satisfactory.
[][]If the equipment found out of calibration limit affix “UNDER Maintenance” label and inform the respective Supplier.

Operating procedure:

[][]Check the machine is clean & calibrated.
[][]Make the switch ON of the viscometer.
[][]Let the machine initialize for 10 minutes before performing Auto zero.
[][]Attach the spindle with the lower shaft of the viscometer.
[][]Set the spindle no. by pressing SELECT SPINDLE button.
[][]Set speed by pressing SET SPEED button.
[][]Select the data parameters by pressing SELECT DISPLAY button.
[][]Insert and center spindle in the test materials until the fluid’s level is at the immersion groove on the spindle‘s shaft.
[][]Press the motor ON/OFF/ESCAPE key and turn the motor ‘OFF’ when changing a spindle or changing samples.

Cleaning procedure:

[][]After completion of the testing switch ‘OFF’ the instrument and switch ‘OFF’ the mains.
[][]Remove the spindle from the instrument.
[][]Clean the spindle with respective solvent in which solution of sample is prepared.
[][]Clean the spindle with methanol.
[][]Wipe it with tissue paper.
[][]Place the clean spindle at its respective place in the box.

Annexure:

Annexure-I: Calibration Information Sheet for Digital Viscometer
Annexure-II: Calculation Sheet for Viscometer Calibration
Annexure-III: Range for LV Spindle cP (m PaºS )

Calibration of Viscometer With Operation And Cleaning Read More »

Calibration of pH meter with operation and cleaning

Calibration of pH meter, Purpose :

Calibration of pH meter, The purpose of this SOP is to describe the operation, calibration and cleaning of pH meter (Model: Mettler Toledo Seven Compact S220) used for the measurement of pH in the quality control and microbiology laboratory.

Calibration of pH meter, Scope :

This procedure describes the application of pH meter, for the determination of pH, in the analysis of raw materials, intermediates and formulated drug product in the quality control and microbiology laboratory of xx Pharmaceuticals Ltd.

Definitions / Abbreviation:

[][]SOP: Standard Operating Procedure.
[][]QC: Quality Control.
[][]pH: The pH is a number which represents conventionally the hydrogen ion concentration of an aqueous solution. It is the measurement of the acidity or alkalinity of an aqueous solution.
[][]Calibration: The set of operations which establish, under specified conditions, the relationship between values indicated by a measuring instrument or measuring system or values represented by a material measure and the corresponding known values of a reference standard.
[][]KCl: Potassium chloride.

Procedure:

Precaution(s):

[][]Store the electrode in 3M KCl solution; never store it in distilled or demonized water or any solution containing heavy metal ions as it may block the electrode membrane.
[][]Always keep the pH electrode wet and rinse it thoroughly with the sample to be measured before use.
[][]Avoid high humidity environments and static discharges since these are detrimental for a stable pH reading.
[][]Do not measure pH of a solution with high temperature.
[][]Try to avoid special measuring solutions such as Sulphide, high Alkaline, Albuminous solutions those can reduce the life of electrode.
[][]Never stir the electrode vigorously in solution during measurement of pH.
[][]After washing, dry the electrode with tissue paper but do not wipe the end of with tissue paper.
[][]Never stir the electrode vigorously in solution during measurement of pH.
[][]After washing, dry the electrode with tissue paper but do not wipe the end of with tissue paper.
[][]Always use fresh and unused buffer for calibration.
[][]Perform a recalibration:
[][]When reference electrode is changed.
[][]After testing aggressive chemicals (such as strong acid or base).

Calibration:

[][]Switch on the power from its main.
[][]Power on the pH meter and printer.
[][]Remove the electrode from the potassium chloride (KCl) solution, rinse it with purified water and soak the excess water with tissue paper very carefully.
[][]Calibrate the instrument daily before use by the three buffer solutions of pH values 7.00, 4.00 and 10.0.
[][]Immerse the electrode in buffer solution of pH value of 7.00 and stir gently and press [CAL] .
[][]Wait until a stable endpoint has been reached.
[][]Hear the ‘beep’ point.
[][]See the sign (/A) at the right side of the display.
[][]Rinse the electrode with purified water and soak the excess water with issue paper very carefully.
[][]Repeat the steps from 7.1.5 to 7.1.9 for second and third calibration for the standard solution of pH values respectively 4.00 and 10.00.
[][]Select “End”.
[][]Select “Save”.
[][]Select “Data”.
[][]Select “Calibration Data”.
[][]Select “pH”.
[][]Select “Review”.
[][]Select “Expertpro ISM”.
[][]Select “Transfer”.
[][]Select “Exit” for 5 times to reach the main menu.
[][]The slope should be 90% – 105%, i.e. at print data, S(B1/B2) and S(B2/B3) = 90% – 105% and offset value will be ± 30 mV, i.e. at print data, off.(B1/B2) and off.(B2/B3) = ±30 mV.

Electrode condition:

DisplaySlope value (%)Condition
Display Icon90 - 105Electrode is in good condition.
Display Icon85 - 89Electrode needs cleaning.
Display Icon84 - 80Electrode is faulty.

Maintenance of the Reference Electrode:

[][]If the electrode slope value falls rapidly, or if the response becomes sluggish or inaccurate, the following procedure should be performed one by one.
[][]Change the reference electrolyte solution every six months.
[][]Degrease the membrane with cotton wool soaked in soap solution.
[][]Soak the tip of the electrode in 0.1M hydrochloric acid overnight.
[][]If all these fail, the electrode must be replaced with a new electrode.
[][]Any replacement of the electrode must be recorded in the log book
[][]Verify the pH of any check buffer (buffer range: 4-9 pH), if reading is within limit (Certificate value ± 0.05), calibration confirms.
[][]Remove the electrode from buffer solution, rinse it with purified water and soak the excess water with tissue paper.
[][]Rinse the electrode with saturated potassium chloride (KCl).
[][]7.1.26 Record the calibration status in Logbook for daily calibration of pH Meter (Seven Compact S220) as per Annexure – I
[][]Power off the pH meter and printer.
[][]Switch off the power from its main.

Operation

[][]Ensure that working area is clean.
[][]Pour enough sample solution into a measuring beaker so that the level of the sample is above the junction of the electrode.
[][]Ensure that the pH meter is cleaned and calibrated on the day use.
[][]Remove the electrode from the potassium chloride (KCl) solution, rinse it with purified water and soak the excess water with tissue paper very carefully. .
[][]Switch on the power its main.
[][]Power on the pH meter and printer.
[][]Gently stir the sample and dip the pH electrode into the solution.
[][]Press [READ] button and wait until a stable endpoint has been reached.
[][]Hear the “beep” sound.
[][]See the sign (/A) at the right side of the display.
[][]Select “Data”.
[][]Select “Measurement Data”.
[][]Select “Review”.
[][]Select “Partial” by using “↓”.
[][]Select “by measurement mode” by using “↓”.
[][]Select “pH”.
[][]Select “transfer”.
[][]Collect the printed result from the printer.
[][]Select “Exit” for 6 (six) times to reach the main display/menu.
[][]Take the electrode out of the solution and rinse with purified water and soak the excess water by tissue paper.
[][]For additional samples repeat the steps from 7.2.7 to 7.2.20 until the sample has been measured.
[][]After taking the measurement, rinse the electrode with saturated potassium chloride (KCl).
[][]Record the measurement data in the Operation Logbook of pH Meter (Seven Compact S220) as per Annexure – II.
[][]Power off the pH meter and printer.
[][]Switch off the power from its main.

Cleaning and maintenance:

[][]Clean the outside case of the pH meter using a mild detergent or disinfectant (e.g.70% ethanol) if required.
[][]Clean the area around the pH meter and wipe away any spillage immediately. Special care should be exercised when cleaning up concentrated HCl (usually) or concentrated NaOH (usually) used for adjusting the pH of solutions.
[][]When required, clean all wires and contacts to avoid corrosion.
[][]Inspect the cables of the electrodes for any signs of broken insulation.
[][]Inspect the pH electrode for any cracks in its stem or bulb.
[][]Rinse off any salt deposits on the top of the electrode using water.
[][]If contamination of the glass bulb is suspected, restoration can be attempted as follows:
[][]Soak electrode in 0.1 M HCl or 0.1 N HNO3 for 15 minutes.
[][]Oily or greasy deposits may be removed by methanol.
[][]After the cleaning stage, rinse the electrode in distilled water and replace the internal filling solution (4M KCl if appropriate) for at least 1 hour before taking any measurements.
[][]Make sure the electrode is always kept filled with the appropriate filling solution.

Annexure: Calibration of pH meter

Annexure-I: Logbook for daily calibration of Seven Compact S220 pH Meter.
Annexure-II: Operation Logbook of Seven Compact S220 pH Meter.

Calibration of pH meter with operation and cleaning Read More »

Calibration of Conductivity Meter with Operation and Cleaning

Calibration of Conductivity, Purpose:

Calibration of Conductivity, The purpose of this SOP is to describe the operation, calibration and cleaning of Conductivity meter (Model: Mettler Toledo Seven Compact S-230) used for the measurement of conductivity of water and any other liquid/solution in the quality control laboratory.

Calibration of Conductivity Meter Scope:

This procedure describes the application of Conductivity meter used for the measurement of conductivity of water any other liquid/solution in the quality control laboratory of XX Pharmaceuticals Ltd.

Definitions/Abbreviation:

[][]SOP: Standard Operating Procedure
[][]QC: Quality Control
[][]Conductivity: Conductivity is the ability of a material to conduct electric current.

Responsibilities:

Officer/Executive/Sr. Executive, Quality Control

[][]To ensure that this procedure is followed.
[][]To maintain the records properly as per SOP.

Manager, Quality Control

[][]To ensure that this procedure is kept up to date.
[][]To confirm that the SOP is technically sound and reflects the required working practices.
[][]To arrange training on the SOP to all concerned personnel and to ensure implementation of the SOP after training.

Head of Quality Assurance

[][]Approval of the SOP.
[][]To ensure the overall implementation of the SOP.
[][]Schedule calibration of the instrument at the defined intervals.

Procedure:

Precaution(s):

[][]Never unscrew the two halves of the housing.
[][]Any spillage should be wiped off immediately some solvents might cause corrosion of the housing.
[][]Avoid the following environment influences:
[][]Powerful vibrations
[][]Direct sunlight
[][]Atmospheric humidity greater than 80%
[][]Corrosive gas atmosphere
[][]Temperature below 5˚C and above 40˚C
[][]Powerful electric or magnetic fields.

Calibration:

[][]Calibrate the instrument on the day of use.
[][]Ensure that working area is clean.
[][]Power on from its main.
[][]Switch on the instrument and printer.
[][]Press “Menu”.
[][]Select “Calibration setting”, then “Calibration standard”.
[][]Select “Predefined standard”, then “Predefined international”.
[][]Select “Supplied standard value of 84 µs/cm, then press “Exit” 2 times return menu.
[][]Take standard into a clean beaker.
[][]Place the electrode into the standard solution.
[][]Press [CAL] wait until a stable endpoint has been reached.
[][]Wait Hear the “beep” sound.
[][]See the sign (/A) at the right side of the display.
[][]Select “Save”, then Select “Data”.
[][]Select “Calibration Data”, then select “Review”.
[][]Select “In Lab 741”, and then select “Transfer” to print the result.

Operation: Calibration of Conductivity Meter

[][]Ensure that working area is clean.
[][]Pour sample into a clean beaker.
[][]Switch on the power its main.
[][]Power on the conductivity meter and printer.
[][]Place the electrode into the sample.
[][]Gently stir the sample and dip the conductivity electrode into the solution.
[][]Press wait until a stable endpoint has been reached.
[][]Hear the “beep” sound.
[][]See the sign (/A) at the right side of the display.
[][]Select “Data”, then “Measurement Data”.
[][]Select “Review”, then “Partial” by using “↓”.
[][]Select “transfer”.
[][]Collect the printed result from the printer.
[][]Select “Exit” for 5 (Five) times to reach the main display/menu.
[][]Record the measurement data in the Operation Logbook of Conductivity meter as per Annexure-II.
[][]Power off the conductivity meter and printer.
[][]Switch off the power from its main.

Cleaning and maintenance:

[][]Wipe the conductivity meter with damp cloth. The housing of the instrument may attack by some organic solvents, such as toluene, xylene and methyl ethyl ketone.

[][]Wipe away any spillage immediately.

Annexure:

Annexure-I: Logbook for Daily Calibration of Conductivity Meter.
Annexure-II: Operation Logbook of Conductivity Meter.

Calibration of Conductivity Meter with Operation and Cleaning Read More »

Calibration of Dissolution Tester with Operation and Cleaning

Calibration of Dissolution Tester Purpose :

Calibration of Dissolution Tester; The purpose of this SOP is to describe the operation, calibration and cleaning of Dissolution Tester (Model: Electrolab, TDT-08L).

Scope :

This procedure is applicable for Dissolution Tester (Model: Electrolab, TDT-08L), installed in the quality control laboratory at general block of XX Pharmaceuticals Limited.

Definitions/Abbreviation:

[][]SOP: Standard Operating Procedure
[][]QC: Quality Control
[][]Dissolution test: Dissolution test is a means to monitor the rate of release of a drug substance from a dosage form to ensure consistency of manufacture and compliance with release specifications.
[][]RPM: Rotations per minute
[][]LED: Light emitting diode
[][]Concentricity: The quality of having the same center (as circles inside one another)
[][]Wobble: Move sideways or in an unsteady way
[][]NMT: Not more than

Responsibilities:

The roles and responsibility is as follows:

Officer/Executive/Sr. Executive, Quality Control

[][]To ensure that the instructions of this procedure are correctly followed.
[][]To ensure cleaning of dissolution tester maintaining safety rules.

Manager, Quality Control

[][]To ensure that this procedure is kept up to date.
[][]To confirm that the SOP is technically sound and reflects the required working practices.
[][]To arrange training on the SOP to all concerned personnel and to ensure implementation of the SOP after training.
[][]Schedule calibration of the instrument at the defined intervals.

Head of Quality Assurance

[][]Approval of the SOP
[][]To ensure the overall implementation of the SOP.

Procedure:

[][]General precautions or operational safety:
[][]Do not start the heater if there is no water in the tank.
[][]Always maintain the liquid level within the tank above the minimum level mark using purified water only.
[][]Disconnect the power supply before moving, emptying or cleaning of bath.
[][]Spillage must be cleaned up immediately after use.
[][]Do not use any aggressive material or strong solvent to clean the jar and water bath.
[][]Locate the system on a level bench without movement or distortion.
[][]Do not hold the stirrer while in operation.
[][]For lifting up the stirrer, make sure the paddles are stop.
[][]Do not disturb the sensor tube while cleaning the tank.
[][]External probe should be handled with care.
[][]Use sinkers if the dosage form unit floats. Close the vessels with the cover during Operation.

Operation:

[][]Connect the instrument to the main power supply.
[][]Ensure the calibration status is valid.
[][]Lift the head by approximately 20 cm by pressing lift UP arrow push button.
[][]Fill the purified water in the tank up to the mark level without jars. Place the tank on the base of the instrument such that four screws at the bottom of the tank fits in to the holes on the base plate and heater cable is at the rear side of the instrument.
[][]Connect the heater plug in the socket provided at the right hand side of the stirrer unit. Add 0.01% Cetyl puridinium chloride or Sodium benzoate or Bronidiol to prevent fungal growth in the water bath.
[][]Connect the power cord from the ETC-11L to TDT-08L in the socket provided on the rear side of the instrument.
[][]Turn on the red colored power switch provided on the front side of ETC-11L.
[][]The yellow colored switch of the Heater will remain always on.
[][]Switch on the power switch provided on the rear

ELECTROLAB
TDT-08L VER 3.50C
DISSOLUTION
TEST APPARATUS

[][]After the power flash screen an idle screen will be displayed showing the last Mode. Protocol no. selected, RPM, Bath temperature and Temp by external probe.
[][]There are 20 protocols. We can assign 20 types of samples against these protocols.
[][]Set the protocol with // keys.
[][]Insert the test vessels in the water bath.
[][]Insert the paddles/basket in the stirrer unit.
[][]Press TEMP key from the front panel. Temperature setting screen will be displayed.
[][]Using // key set the temperature. Set the temperature to 37.7ºC so that the jar temperature reaches to 37ºC.
[][]Press F1 to on the heater. The [Temp. ON] indicator will flash on.
[][]Press ENTER key to register the temperature and come out of the TEMP mode. An idle screen will be displayed.
[][]Press RPM key. Adjust the RPM with // keys.
[][]Press ENTER to register the RPM and come out of the screen.
[][]To set the time press TIME key. A screen will appear like

SAMP. No = 00 OF 00 
SAMP. INTV. = 00 Hr: 00 Mn

[][]Set the sampling number 01 OF 01. Set total dissolution time (as per test method for a particular product) in SAMP. INTV by using // keys.
[][]Press ENTER key to register the time and come out to the initial screen.
[][]Wait to raise the bath temperature up to the desired level.
[][]When the temperature will raise to desired level the [READY] indicator will flash on.
[][]Add dissolution media into individual test vessel as per test method. Measure media temperature with the sensor. Wait until the temperature reaches to 37 ± 0.5°C.
[][]Place sample in vessels in case of paddle apparatus taking care to exclude air bubbles from the surface of the dosage form unit.
[][]In case of basket apparatus place sample in dry baskets, attach the shaft into stirrer unit and lower down the stirrer unit into medium. By lift DOWN key.
[][]Immediately press START key to run the operation.
[][]The screen will show the Elapse time of operation.
[][]After completion of the time interval the instrument will give beep sound.
[][]Lift the stirrer using the lift UP key.
[][]Withdraw the dissolution medium with suitable pipette from a zone midway between the surface of the dissolution medium and the top of rotating paddle through the head nozzle, not less than 1 cm from the vessel wall.
[][]The sampling has to be filtered immediately after completion of the sampling.
[][]After completion of work, switch ‘OFF’ the instrument.
[][]Switch ‘OFF’ the mains when not required.
[][]Fill up the operation log book for Dissolution tester as per Annexure-III.

Physical Calibration

Frequency: Perform mechanical calibration of the dissolution tester once in every six months or if the system has been moved, serviced or a malfunction is suspected.

For Temperature Calibration:

[][]Switch on the mains
[][]Set the temperature to 37°C by following the temperature setting procedure given in operation.
[][]Ensure that the jars are filled with water
[][]Allow the instrument to stabilize
[][]Measure the temperature of the water bath and of each jar with a calibrated thermometer and compare the result against the digital display on the apparatus.
[][]Record the observation as per Annexure-I.
[][]Acceptance Criteria: 37°C ± 0.5°C.

For RPM Calibration:

[][]Switch on the mains.
[][]Set the RPM to 50 as per procedure given in the operation.
[][]Start the stirrer and measure the RPM by using calibrated tachometer.
[][]Record the observation as per Annexure-I.
[][]Similarly check the RPM 75, 100, and 150 and record the observations.
[][]Acceptable criteria: ± 4% of set value.

For Time Calibration:

[][]Switch on the mains.
[][]Set the desired time (30 and 45 Minutes) by following the temperature setting procedure given in operation.
[][]Check the time using a calibrated stop watch and record the observations as per Annexure-I.
[][]Acceptable criteria: ± 2% of set value.

Chemical Calibration (Dissolution Performance Verification Testing)

Frequency: Perform chemical calibration of the dissolution tester once in a year or if the system has been moved, serviced or a malfunction is suspected.

For USP Dissolution Apparatus 1 (basket) and Apparatus 2 (paddle):

[][]If the instrument is outside calibration, a label of “Out of calibration” is placed.
[][]Apparatus must be calibrated with the following :

Disintegrating Type – USP Prednisone Tablets 10 mg at 50 rpm

[][]Non-disintegrating Type – USP Salicylic Acid Tablets 300 mg at 100 rpm
[][]Checking Standards in the on line-USP and printing Lot-Specific Standard Certificates. Verifies Standard Lots are current within expiry.
[][]Disintegrating Type – USP Prednisone tablets 10 mg at 50 rpm:
[][]Follow the directions for storage and use on the label of the reference standard.
[][]If drying instructions are included on the label, only dry sufficient quantity required for the test (use a clean, dry vessel when drying and not the original container).
[][]Standard solutions must be prepared on the day of use.

Prednisone Standard Preparation (Standards must be prepared in duplicate):

[][]Accurately weigh 20 mg Prednisone standard into a 100 ml volumetric flask containing 5 ml of methanol.
[][]Sonicate to dissolve and dilute to volume with distilled or purified water and mix well.
[][]Further dilute 5.0 ml of this solution to 100 ml with distilled or purified water and mix well.

Dissolution Medium Preparation and De-aeration:

[][]Filter the dissolution medium (Distilled or purified water 500 ml) with the aid of a vacuum through a 0.45 µm-porosity membrane filter into a suitable filtering flask.

Test Procedure:

[][]Using a volumetric flask, measure 500 ml distilled or purified water into each of the six vessels.
[][]Allow medium to equilibrate for approximately 30 minutes before test is commenced.
[][]When dissolution bath has equilibrated, check the temperature of each vessel is 37.0°C ± 0.5°C.
[][]Carefully weigh 6 de-dusted USP Prednisone Tablets 10 mg.
[][]Drop one previously de-dusted tablet into each of the vessels via the sample holes in the lids, raising the stirrer shaft fully before dropping the tablet into the vessel, and pushing the shaft fully down when the tablet has reached the bottom.
[][]Operate the apparatus at 50 rpm for 30 minutes.
[][]After 30 minutes withdraw approximately 50 ml midway between the surface of the dissolution medium and the top of the blade, not less than 1cm from the vessel wall.
[][]Filter through a Whatman Number 1 filter paper, discarding the first few ml of filtrate.
[][]Scan the standard from 300-200 nm to obtain the maximum absorbance at the optimum wavelength (approximately 242 nm).
[][]Determine the absorbance of the six samples at that wavelength in fixed mode.
[][]Calculate the percentage prednisone dissolved for all 8 stations using the following calculations:
Absorbance of Sample x Standard Concentration (mg/l) x Standard Potency x 100] divided by
[Absorbance of Standard x Sample Concentration if 100% dissolved (C100) x 100]

Acceptance Criteria:

[][]Check the percentage released limit of USP dissolution calibration tablets (Disintegration type: Prednisone tablets) from the respective batch certificate provided by USP.
[][]If any one station is outside of specification, place an “Instrument out of calibration, DO NOT USE” label on the instrument.
[][]Inform In-charge, QC; if the results do not meet the criteria.

Test interpretation:

[][]The apparatus shall be suitable if the individual calculated values at each indicated speed are within the range specified in the tablet obtained with the each lot of tablets.
[][]Note: An amount of alcohol not to exceed 5% of the total volume of the standard solution may be used to bring the prednisone standard into solution prior to dilution with dissolution medium.

Recording of Results:

[][]All relevant information on the calibration must be completed on Annexure-II.
[][]When all the results have been checked affix a Calibration label to the instrument.

Cleaning procedure

[][]Lift the stirrer unit by using “lift UP” key.
[][]Switch ‘OFF’ the heater.
[][]Switch ‘OFF’ the instrument.
[][]Switch ‘OFF’ the mains.
[][]Remove the plug from the main socket.
[][]De-dust the outer surface of the instrument with a clean dry cloth every day.
[][]Clean apparatus after completion of every dissolution test.
[][]Remove the test vessels from their place and discard the medium.
[][]Wash each vessel with purified water and then set them inside the dissolution bath number wise.
[][]Remove basket/paddle apparatus from the dissolution tester. Wash them with purified water and dry with tissue paper and keep at dedicated place.
[][]Clean and replace the water from dissolution bath fortnightly (once in every two weeks) or if necessary.
[][]Remove all vessels from the dissolution bath.
[][]Remove the basket/ paddle apparatus from the stirrer unit.
[][]Remove the circulating pipe which is provided at the right rare side of the dissolution bath.
[][]Start the circulation pump from the option menu and drain the dirty water in a waste water bucket.
[][]Insert the circulating pump in its proper place.
[][]Wash the dissolution bath with potable water, detergent solution and then again with potable water sequentially until no foam of detergent remains in the bath.
[][]Wash the bath finally with purified water and fill the bath with the same up to the level mark.

 

Annexure Download Here

Annexure-I: Physical Calibration Information Sheet for Dissolution Teste
Annexure-II: Chemical Calibration Information Sheet for Dissolution Tester
Annexure-III: Operation Log book for Dissolution Tester

 

Calibration of Dissolution Tester with Operation and Cleaning Read More »

Calibration of UV-Vis Spectrophotometer With Operation & Cleaning

Calibration of UV-Vis Spectrophotometer Purpose :

Calibration of UV-Vis Spectrophotometer, The purpose of this SOP is to describe the operation, calibration and cleaning of UV-Vis Spectrophotometer (Shimadzu, UV-1800) used in the quality control laboratory of General block at XX Pharmaceuticals Limited.

Scope :

This procedure is applicable for UV-Vis Spectrophotometer (Shimadzu, UV-1800), installed in the quality control laboratory of XX Pharmaceuticals Limited.

Definitions / Abbreviation:

[][]SOP: Standard Operating Procedure.
[][]QC: Quality Control.
[][]UV-Vis: Ultraviolet and Visible.

Responsibilities:

Sr. Executive/Executive, QC

[][]To ensure that this procedure is followed.
[][]To maintain the records properly as per SOP.

Manager, Quality Control

[][]To ensure that this procedure is kept up to date.
[][]To confirm that the SOP is technically sound and reflects the required working practices.
[][]To arrange training on the SOP to all concerned personnel and to ensure implementation of the SOP after training.
[][]Schedule calibration of the instrument at the defined intervals.

Head of Quality Assurance

[][]To ensure the overall implementation of the SOP.
[][]Approval of the SOP

Procedure:

Precaution(s):

[][]Prior to use, user must ensure that equipment is calibrated.
[][]Handle the measuring cell with care.
[][]Do not forget to turn off the instrument after use or if the next reading will be taken after long period of time.

Operation: Calibration of UV-Vis Spectrophotometer

[][]Switch ‘ON’ the Computer, Printer and Spectrophotometer.
[][]Wait until initialization is completed and the initialization result will be displayed in the outlet screen of the machine, finally green signal will appear in the outlet light-bar.
[][]Double click on ‘UV probe’ icon in the computer to run the software.
[][]Press ‘Connect’ icon to establish communications with the machine and machine will be ready for any type of analysis in Photometric/ Spectrum mode.

Measurement in spectrum mode

[][]Click “Spectrum” icon to perform measurement in spectrum mode.
[][]Click “Method” icon to set the desired parameters according to the requirement as follows:
[][]Click ‘Measurement’ icon then set following steps:
[][]Wavelength Range: Start & End- as required.
[][]Scan Speed : Fast/ Medium/ Slow/ Very slow
[][]Sampling Interval : As required
[][]Scan mode: Single/ Auto
[][]Input desired filename for the respective data to preserve in distinguishable data-path.
[][]Click ‘Instrument parameter’ to set the measuring mode as: Absorbance or transmittance.
[][]Finally press “OK” to resolve the setup, click ‘Close’.
[][]Click ‘ Baseline’ icon followed by ‘OK’ for baseline correction.
[][]Rinse both the cells with respective blank solution & wipe properly the outer surface of the cell with tissue paper. Put the cells in cell holder and close the curvature.

[][]Click “Auto zero” to compensate the blank reading.
[][]Remove one cell (front side) and after proper rinsing with the required sample solutions fill the cell with sample solution, wipe properly the outer surface of the cell with tissue paper & Set sample in the cell holder and finally close the curvature.
[][]Click “START”. The photometer status window will show “Slewing….” and it will be followed by the active reading.
[][]After completion of measurement click ‘Point pick’ for determining the absorbance/ transmittance at definite wavelength or Click ‘Peak pick’ for determining the absorbance/ transmittance at a given wavelength range.
[][]Click ‘Save’ to preserve the data and print the data on respective report format.
[][]Wash the measuring cells properly with purified water and preserve in appropriate place.

Measurement in quantitative mode

[][]Click “Photometric” icon to perform measurement in quantitative mode.
[][]Click “Method” icon to set the desired parameters according to the requirement as follows:
[][]Wavelength type: point
[][]Input required wavelength in ‘Wavelength (nm)’, click ‘Add’ then press Next
[][]Input ‘Type’ as ‘Single point’, Formula as ‘Fixed Wavelength’. In ‘WL1’ activate the required wavelength, then input standard concentration in ‘STD Concentration’, click Next>Next>Next.
[][]Finally in ‘Photometric method’ input ‘filename’ for the respective data to preserve in distinguishable data-path. Also input ‘Title’ if any, user name in ‘Analyst’ and fill-up ‘Comments’ if any.
[][]Press “Finish” to resolve the setup, click ‘Close’.
[][]Rinse both the cells with respective blank solution &swipe properly the outer surface of the cell with tissue paper. Put the cells in cell holder and close the curvature.

[][]Click “Autozero” to compensate the blank reading.

[][]Remove one cell (front side) and after proper rinsing with the required standard solutions fill the cell with standard solution, wipe properly the outer surface of the cell with tissue paper & Set standard solution in the cell holder and finally close the curvature.
[][]In the ‘Standard table’ input the standard description in the column for ‘Sample ID’, Click on ‘WL column’ and finally click ‘Read Std’ icon to measure the result.
[][]Remove the standard solution cell (front side) and after proper rinsing with the required sample solutions fill the cell with sample solution, wipe properly the outer surface of the cell with tissue paper & Set sample in the cell holder and finally close the curvature.
[][]In the ‘Sample table’ input the sample description in the column for ‘Sample ID’, Click on ‘WL column’ and finally click ‘Read Unk.’ icon to measure the result.
[][]Click ‘Save’ to preserve the data and print the data on respective report format.
[][]Wash the measuring cells properly with purified water and preserve in appropriate place.
[][]Click ‘Disconnect’ icon to detach the software from the machine.
[][]Exit from the ‘UV probe’ software by closing the window. Put off the power of equipment, printer and computer.

Calibration of UV-Vis Spectrophotometer:

[][]Calibrate the UV-Vis Spectrophotometer at 6 month’s frequency either by following procedure or as per supplier’s protocol:

Wave length Scanning

[][]Set desired start wavelength i.e. higher end of the desired spectrum and desired wavelength expansion. Scan the desired wavelength.

Control of Wavelengths

[][]Verify the wavelength at 241.15 nm, 287.15 nm, 361.5 nm & 536.3 nm using the absorption maximum of Holmium perchlorate solution (4% w/v solution of holmium oxide in a solution of perchloric acid containing 14.1% w/v of HClO4). The permitted tolerance is ± 1 nm for the ultraviolet range and ± 3 nm for the visible range. Record data in calibration information sheet as per Annexure-I .

Control of Absorbance

[][]Check the absorbance using a solution of potassium dichromate at the wavelengths indicated in (Annexure-I) which gives for each wavelength the exact values and the permitted limits of the specific absorbance. The tolerance for the absorbance is ± 0.01.
[][]For the control of absorbance, use solutions of potassium dichromate which has previously been dried to a constant mass at 130° C. For the control of absorbance at 235 nm, 257 nm, 313 nm and 350 nm, dissolve 57.0-63.0 mg of potassium dichromate in 0.005 M Sulphuric acid and dilute to 1000.0 ml with the same acid. For the control of absorbance at 430 nm, dissolve 57.0-63.0 mg of potassium dichromate in 0.005 M sulphuric acid and dilute to 100.0 ml with the same acid.
[][]Measure at a path length of 1 cm, at the wavelength 235 nm, 257 nm, 313 nm, 350 nm and 430 nm against 0.005 M Sulphuric acid as blank. Record data in calibration information sheet as per Annexure-I.

Limit of Stray light

[][]Detect stray light at a given wavelength with suitable solutions: for example the absorbance of a 12 g/l solution of potassium chloride in a 1 cm cell increases steeply between 220 nm and 200 nm and is greater than 2.0 at 198 nm when compared with water as compensation liquid. Record data in calibration information sheet as per Annexure-II.

Resolution (for qualitative analysis)

[][]Prepare a 0.02% v/v solution of toluene in hexane. Use hexane as a blank. The minimum ratio of the absorbance at the maximum at 269 nm to that at the minimum at 266 nm is not less than 1.5 unless otherwise stated in the monograph. Suitable certified reference materials may also be used. Record data in calibration information sheet as per Annexure-II.

Path length measurement of cell

[][]The tolerance on the path length of the cells used is ± 0.005 cm. When filled with the same solvent, the cells intended to contain the solution to be examined and the compensation liquid must have the same transmittance. If this is not the case, an appropriate correction must be applied.
The cells must be cleaned and handled with care.
[][]Carryout the test once before first use of a new cell.
[][]Affix the label “CALIBRATED “on the equipment if all results are satisfactory. If the equipment found out of calibration limit affix “NOT FIT FOR USE” label and call Supplier.

Cleaning procedure: Calibration of UV-Vis Spectrophotometer

[][]After completion of each analysis, clean the UV-Vis spectrophotometer as follows:
[][]Remove cuvette from the sample compartment.
[][]Remove the solution from the cuvette and wash with respective solvent in which solution is prepared.
[][]Rinse with water and then with methanol and wipe with tissue paper.
[][]Air dry and keep it at dedicated place.
[][]When required, clean the outer surface of the instrument with Isopropyl alcohol and dry it with tissue
paper.

Annexure:

Annexure-I: Calibration Information Sheet for UV-Vis Spectrophotometer (Control of Wavelength and Control of Absorbance)
Annexure-II: Calibration Information Sheet for UV-Vis Spectrophotometer (Resolution, Limit of Stray light)
Annexure-III: Operation Logbook for UV-Vis Spectrophotometer

 

 

 

Calibration of UV-Vis Spectrophotometer With Operation & Cleaning Read More »

Calibration of FTIR Spectrophotometer with Cleaning and Operation

Calibration of FTIR, Purpose:

Calibration of FTIR, The purpose of this SOP is to describe the operation, calibration and cleaning of FTIR Spectrophotometer (Shimadzu, IR Affinity-1) used in the quality control laboratory at XX Pharmaceuticals Limited.

Scope

This procedure is applicable for FTIR Spectrophotometer (Shimadzu, IR Affinity-1), installed in the quality control laboratory of XX Pharmaceuticals Limited.

Definitions / Abbreviation:

Standard Operating Procedure (SOP): Standard Operating Procedure. A written authorized procedure, which gives instructions for performing operations.

[][]QC: Quality Control.

[][]FTIR: Fourier Transform Infrared.

Responsibilities

[][]Executive/ Sr. Executive, QC

[][]To ensure that this procedure is followed.

[][]To maintain the records properly as per SOP

Manager, Quality Control

[][]To ensure that this procedure is kept up to date.
[][]To confirm that the SOP is technically sound and reflects the required working practices.
[][]To arrange training on the SOP to all concerned personnel and to ensure implementation of the SOP after training.
[][]Schedule calibration of the instrument at the defined intervals.

Head of Quality Assurance

[][]Approval of the SOP.
[][]To ensure the overall implementation of the SOP.

Procedure: Calibration of FTIR

General Procedure:

[][]Ensure that the instrument is clean and calibrated.
[][]Connect the instrument properly with the power supply.
[][]Switch “ON” the main switch, power switch of the instrument situated at the backside panel, printer and computer.
[][]Allow the instrument initialize for at least 1 hour.

Disk preparation for solid sample:

[][]Mix 5 to 10 mg of sample with 200 to 400 mg of Potassium Bromide (KBr) [IR Grade].
[][]Grind the mixture to very fine powder with agate mortar.
[][]Pour the mixture into the barrel and carefully insert the plunger with its polished face downward.
[][]Maintain the hydraulic pressure 50 KN to 60 KN of the barrel for 1-2 minutes.
[][]Release the pressure and take out the disk from the barrel.
[][]For sample of liquid material using sealed liquid cell:
[][]Inject the liquid sample (pure or in solution) in sealed liquid cell.
[][]Plug to close both the slots of sealed liquid cell.
[][]For sample of liquid material which converted to solid (Nujol preparation) and measurement using NaCl demountable liquid cell:
[][]Put the cell window on the cell window holder and place the spacer of appropriate thickness on the cell window.
[][]Take 3 to 5 mg of sample with 2 to 3 drop of Nujol in a mortar and mix properly.
[][]Insert sample inside the spacer with a spatula. Put another cell window on the spacer and fit the metal plate on it.
[][]Fasten those plates with the four screws. Do not fix them with excessive force. It is enough to fasten them tightly by hand.

Operation Procedure (FTIR): calibration of FTIR

[][]Double click on “IRsolution” software icon.
[][]Press Measure
[][]Select FTIR Parameters in parameter window as follows:
[][]Data :
[][]Measurement Mode: % Transmitance ,
[][]Resolution: 2
[][]Apodization: Happ-Genzel
[][]No of Scans: 10
[][]Range (cm-1): Min: 400 cm-1 Max: 4000 cm-1
[][]Instrument:
[][]Beam: Internal
[][]Detector: Standard
[][]Mirror speed: 2.8
[][]More:
[][]Mode: Power
[][]Remove sample, if any, from the sample compartment.
[][]Click on Measurement then “Initialize” then “Yes”.

[][]After successful initialization, click BKG then “OK”. Wait till window on the screen is displayed as last Bkg.
[][]After ‘Bkg scan’, put the sample disk in the sample chamber, with the help of sample holder and close the chamber.
[][]Enter the Sample name, Batch No., Sample ID. in comment box and enter the data file location in Data file box.
[][]Click Sample to measure sample. Wait till the actual sample IR scan is displayed.
[][]Click on the spectrum window. Select Calculate  peak table. Type threshold: 60.00.
[][]Click on Calc, OK
[][]Compare the Spectrum in computer screen. View the IR scan of standard, click on Manipulation 2 then “Purity” then “Yes”.
[][]Again view the IR scan of sample, click on Manipulation 2  “Purity”  “Calc”  “OK”.
[][]After comparison with standard and sample then click file and open file as required. Then click Window  “Join visible”  “Calc”  “OK”.
[][]Purity index should be within 0 .9500 – 1.0000.
[][]A disc is rejected if visual examination shows lack of uniform transparency or when transmittance at about 2000 cm-1 (5 µm) in the absence of a specific absorption band is less than 60 % without compensation, unless otherwise prescribed in individual monograph.
[][]For printing of spectrum, click File  “Print preview”. Select respective necessary template and then print.
[][]Close the parameter window and the Software window.
[][]Shutdown PC, switch off Printer.
[][]Power OFF Spectrophotometer.
[][]Take out the liquid cell from the instrument.
[][]Discard the disk /Nujol / liquid sample.
[][]Clean the holder and liquid cell with dry and clean tissue paper.
[][]In case of sealed liquid cell mop and wash with IPA or absolute Ethanol and in case of demountable liquid cell mop and wash with Chloroform or Carbon Tetrachloride.
[][]Keep them in specified desiccator.
[][]Rub mortar, pestle, barrel & plunger with tissue paper to make them free from dust.
[][]Mop them with a tissue paper soaked in absolute ethanol.
[][]Keep them in a desiccator for next operation.
[][]Record all analysis in operation log book as per Annexure-II.

Operation Procedure (ATR):

Running a sample:

[][]Turn on the computer and crick on the respective software on the desktop.
[][]Set the MIRacle/ MIRacle A Base Unit to the instrument.
[][]Before initialization, set FTIR Parameters in parameter window as follows:
[][]Data :
[][]Measurement Mode: % Transmitance ,
[][]Resolution:2
[][]Apodization: Happ-Genzel
[][]No of Scans: 10
[][]Range (cm-1): Min : 700 cm-1 Max : 4000 cm-1
[][]Click on Measurement then “Initialize” then “Yes”. Click “cancel” option for auto adjustment.
[][]Click on Measurement then “Auto adjustment (Fine)” then “Yes”.
[][]After autoadjustment, click BKG then “OK”. Wait till window on the screen is displayed as last Bkg.
[][]Set the clamp to the base. Place sample on the ZnSe crystal- Liquids -place a drop of sample on the ZnSe crystal. Place a small neat solid sample on the ZnSe crystal with a spatula. Adjust the knob of the ATR until the pressure reads 12.
[][]Make sure that no acid sample (pH below 5), strong alkaline sample (pH more than 9) or liquid which give damage to PTFE should not be place on ZnSe prism.
[][]Enter the Sample name, Batch No., Sample ID. in comment box and enter the data file location in Data file box. Click Sample to measure sample. Wait till the actual sample ATR scan is displayed.

Working up the Data:

[][]Compare the Spectrum in computer screen. View the ATR scan of standard, click on Manipulation 2 then “Purity” then “Yes”.
[][]Again view the ATR scan of sample, click on Manipulation 2  “Purity”  “Calc”  “OK”.
[][]After comparison with standard and sample then click file and open file as required. Then click Window  “Join visible”  “Calc”  “OK”.
[][]For printing of spectrum, click File  “Print preview”. Select respective necessary template and then print

Calibration:

[][]Calibrate the FTIR Spectrophotometer at 6 month’s frequency either by following procedure or as per supplier’s protocol:
[][]Power ON the Spectrometer. Wait for at least 1 hour. Power ON PC & Printer.
[][]Double click on “IRsolution” software icon. Press Measure
[][]Select FTIR Parameters in parameter window as follows:
[][]Data :
[][]Measurement Mode: % Transmitance ,
[][]Resolution : 2
[][]Apodization : Happ-Genzel
[][]No of Scans : 10
[][]Range (cm-1) : Min : 400 cm-1 Max : 4000 cm-1
[][]Instrument :
[][]Beam : Internal
[][]Detector : Standard
[][]Mirror speed : 2.8
[][]More :
[][]Mode : Power
[][]Remove sample, if any, from the sample compartment.
[][]Click on Measurement then “Initialize” then “Yes”.
[][]After successful initialization, click BKG then “OK”. Wait till window on the screen is displayed as last Bkg.
[][]After ‘Bkg scan’, place the Polystyrene film in the sample chamber and close the chamber.
[][]Enter the Polystyrene 1 and Wave number verification in comment box and enter the data file location in Data file box.
[][]Click Sample to scan the polystyrene film (0.05 mm thick). Wait till the actual sample IR scan is displayed.
[][]Click on the spectrum window. Select Calculate  peak table. Type threshold: 65.00.
[][]Click on Calc  OK
[][]For printing of spectrum, click File  “Print preview”. Select respective necessary template and then print. Collect Polystyrene spectrum and peak table printed in two pages.

Limit of Wave Number Accuracy

[][]Check the maximum at following wave numbers (cm-1): 3060.0 (±1.5) cm-1, 2849.5 (±1.5) cm-1, 1942.9 (±1.5) cm-1, 1601.2 (±1.0) cm-1, 1583.0 (±1.0) cm-1, 1154.5 (±1.0) cm-1, 1028.3 (±1.0) cm-1.
[][]Resolution of apparatus
[][]Resolution due to transmittance:
[][]Record the Spectrum of a polystyrene film of approximately 35 µm thickness using Measuring Mode: % T.
[][]Measure the percentage transmittance at the transmission maximum at 2870 cm-1 (3.48 µm) and that at the transmission minimum at 2849.5 cm-1 (3.51 µm).
[][]Measure the percentage transmittance at the transmission maximum at 1589 cm-1 (6.29 µm) and that at the transmission minimum at 1583 cm-1 (6.32 µm).
[][]Record the results in calibration information sheet as per Annexure-I.

Acceptance criteria:

[][]The difference between the % transmittance at the transmission maximum at 2870 cm-1 (3.48 µm) and that at the transmission minimum at 2849.5 cm-1 (3.51 µm) must be greater than 18.
[][]The difference between the percentage transmittance at the transmission maximum at 1589 cm-1 (6.29 µm) and that at the transmission minimum at 1583 cm-1 (6.32 µm) must be greater than 10.

Resolution due to Absorbance:

[][]Record the Spectrum of a polystyrene film of approximately 35 µm thickness using Measuring Mode: Abs.
[][]Measure the absorbance at the absorption minimum at 2870 cm-1 and the absorption maximum at 2849.5 cm-1.
[][]Measure the absorbance at the absorption minimum at 1589 cm-1 and the absorption maximum at 1583 cm-1.
[][]Record the results in calibration information sheet as per Annexure-I.

Acceptance criteria:

[][]The difference between the absorbance at the absorption minimum at 2870 cm-1 and the absorption maximum at 2849.5 cm-1 is greater than 0.33.
[][]The difference between the absorbance at the absorption minimum at 1589 cm-1 and the absorption maximum at 1583 cm-1 is greater than 0.08.
[][]If the calibration is not within the specified standards, check the cleanliness of polystyrene film again and repeat the calibration.
Wavelength Repeatability Test:
[][]Repeat the Spectrum of the polystyrene film and Superimposed one after another of the absorption bands at 2849.5 cm–1 (3.51 µm), 1601.2 cm–1 (6.25 µm), 1028.3 cm –1 (9.72 µm).
[][]Acceptance Criteria: Compare the difference within 0.5% of the wave number scale and Purity index should be within 0.9500 – 1.0000
[][]Affix the label “CALIBRATED “on the equipment if all results are satisfactory. If the equipment found out of calibration limit affix “NOT FIT FOR USE” label and inform supplier.

Cleaning procedure:

[][]Clean the instrument with clean dry cloth daily.
[][]Remove your sample from the ATR, by using a cloth that is damped with acetone.
[][]Clean the plunger in the same way, if it made contact with sample.
[][]Never squirt (spray) acetone directly on the platform.

Annexure: Calibration of FTIR

Annexure-I: Calibration Information Sheet for FTIR Spectrophotometer
Annexure-II: Operation Logbook for FTIR Spectrophotometer

Calibration of FTIR Spectrophotometer with Cleaning and Operation Read More »